首页    期刊浏览 2025年12月04日 星期四
登录注册

文章基本信息

  • 标题:Determination of raloxifene hydrochloride in human urine by LC-MS-MS
  • 本地全文:下载
  • 作者:U.R.A. Kumar ; K. Basavaiah ; K. Tharpa
  • 期刊名称:Chemical Industry and Chemical Engineering Quarterly
  • 印刷版ISSN:1451-9372
  • 出版年度:2009
  • 卷号:15
  • 期号:03
  • 页码:119-123
  • 出版社:Association of the Chemical Engineers
  • 摘要:A sensitive and selective liquid chromatographic-tandem mass spectrometric (LC-MS-MS) method was developed to determine raloxifene hydrochloride (RLX) in human urine. After a solid-phase extraction with SPE cartridge, the urine sample was analyzed on a C18 column (Symmetry 3.5μm; 50 mm×4.6 mm i.d) interfaced with a triple quadruple tandem mass spectrometer. A positive electrospray ionization was employed as the ionization source. The mobile phase consisted of ammonium acetate (pH 4.0)–acetonitrile (60:40, v/v).The method was linear over a concentration range of 20–1000 ng mL-1. The lower limit of quantitation was 20 ng mL-1. The intra-day and inter-day relative standard deviation across three validation runs over the entire concentration range was <10.5%. The accuracy determined at three concentrations (50, 500 and 850 ng mL-1 RLX) was within ±0.84% in terms of relative errors.
  • 关键词:raloxifene ; quantification ; LC-MS-MS ; urine
国家哲学社会科学文献中心版权所有