  • 标题:A comparison of in-air and in-water calibration of a dosimetry system used for radiation dose assessment in cancer therapy
  • 本地全文:下载
  • 作者:Arshed Waheed ; Mahmood Khalid ; Qazi Ikramullah
  • 期刊名称:Nuclear Technology and Radiation Protection
  • 印刷版ISSN:1451-3994
  • 出版年度:2010
  • 卷号:25
  • 期号:1
  • 页码:51-54
  • DOI:10.2298/NTRP1001051A
  • 出版社:VINČA Institute of Nuclear Sciences
  • 摘要:

    An accurate calibration of the therapy level radiation dosimetry system has a pivotal role in the accuracy of dose delivery to cancer patients. The two methods used for obtaining a tissue equivalent calibration of the system: air kerma calibration and its conversion to a tissue equivalent value (absorbed dose to water) and direct calibration of the system in a water phantom, have been compared for identical irradiation geometry. It was found that the deviation between the two methods remained within a range of 0% to ±1.7% for the PTW UNIDOS dosimetry system. This means that although the recommended method is in-water calibration, under exceptional circumstances, in-air calibration may be used as well.

  • 关键词:absorbed dose calibration; kerma to absorbed dose conversion