Experimental verifications of two optimized adaptive digital signal processing algorithms implemented in two pre set time count rate meters were per formed ac cording to appropriate standards. The random pulse generator realized using a personal computer, was used as an artificial radiation source for preliminary system tests and performance evaluations of the pro posed algorithms. Then measurement results for background radiation levels were obtained. Finally, measurements with a natural radiation source radioisotope 90Sr-90Y, were carried out. Measurement results, con ducted without and with radio isotopes for the specified errors of 10% and 5% showed to agree well with theoretical predictions.