首页    期刊浏览 2024年07月19日 星期五
登录注册

文章基本信息

  • 标题:Quantitative Determination of Di (2-Ethylhexyl) Phthalate (DEHP) in Hemodialysis-Related Materials
  • 本地全文:下载
  • 作者:Mitra Mahdavi Mazdeh ; Gholamreza JahedKhaniki ; Masoud Yonesian
  • 期刊名称:Knowledge & Health Journal
  • 印刷版ISSN:1735-577X
  • 出版年度:2012
  • 卷号:7
  • 期号:2
  • 出版社:Shahroud University of Medical Sciences
  • 摘要:Introduction: Phthalates are founded in medical devices such as filters and dialysis catheters. Scientific evidences show health disadvantages due to exposure to phthalates. In this study, level of Di (2-ethylhexyl) phthalate (DEHP) in Iranian hemodialysis-related materials was measured. Methods: Ten samples of Iranian dialysis catheters (five samples from SUPA medical devices company (SUPA-MDC) and five samples from Helal Ahmar- MDC) were randomly selected. The level of DEHP for each sample was measured by Gas chromatography- Mass spectrometry (GC-MS). Mean concentrations of DEHP (mg per ml) for each brand was reported separately. Results: Means of DEHP concentration for SUPA- MDC and Helal Ahmar- MDC were 1.36±0.11 and 0.97±0.11 mg/ml, respectively. Range of measured concentrations differed from 1.47 to 1.21 mg/ml and 1.13 to 0.83 mg/ml, for SUPA- and Helal Ahmar-MDCs respectively. Conclusion: Application of alternative medical products without or with less phthalate could reduce exposure of patients to phthalates
  • 关键词:Di (2-ethylhexyl) phthalate; Catheters; Hemodialysis