首页    期刊浏览 2024年07月08日 星期一
登录注册

文章基本信息

  • 标题:Removal of Alcyon II, CGR, MeV 60Co Teletherapy Head and Evaluation of Exposure Dose
  • 本地全文:下载
  • 作者:Yasser Tawflk Selim ; Yasser Fouad Lasheen ; Mahmoud Ali Hassan
  • 期刊名称:Journal of Environmental Protection
  • 印刷版ISSN:2152-2197
  • 电子版ISSN:2152-2219
  • 出版年度:2013
  • 卷号:4
  • 期号:12
  • 页码:1435-1440
  • DOI:10.4236/jep.2013.412164
  • 出版社:Scientific Research Publishing
  • 摘要:Teletherapy units containing 60Co source are used in Egypt and worldwide for treatment of cancer diseases. One of the widely used in Egypt is Alcyon CGR MeV. Different radiotherapy facilities need to unload this 60Co unit after being disused. The head containing the source was removed and put into approved package for transportation, after the source kept in safe position. This study describes the planning and removing process of the disused sources and, transports it into its head. The personal doses measured during the process. This work pre-estimates the doses received to workers during the removing process. The workers received doses of approximately 55 μSv during the whole process. This study shows that the original lead head can be used as the source container of this 60Co unit but not as a long term storage option. The 60Co machine was smoothly dismantled and transported by trained workers and driver, using planned and controlled radiation protection, followed by ALARA (as low as reasonably achievable) rule.
  • 关键词:Alcyon; <sup>60</sup>Co; Teletherapy; Head; Sealed Source