摘要:Objective: To accurately measure the end-expiratory pulmonary artery wedge pressure (PAWP) with the “expiration holding” function on the ventilator and the “pulmonary artery wedge pressure review” software on the monitor. Materials and methods: Fifty prospective measurements were made on 12 patients undergoing pulmonary artery catheter and mechanical ventilation. All measurements were divided into <8 mmHg or ≥8 mmHg subgroups according to respiratory variability, and they were then subdivided into either an airway pressure display measurement group (AM group) or an expiration holding (EH) group for comparison. Results: In all measurements, the two groups showed similar levels of accuracy; however, for the time spent for measurement, the EH group was much faster than the airway pressure display measurement group ( P <0.001). Additionally, the EH group was associated with lower medical costs. Conclusion: The expiration holding approach measured the PAWP more accurately, more quickly, and with reduced costs in comparison to the airway pressure display approach.