  • 标题:The effect of infusion rate and catheter length on the temperature of warming fluid
  • 本地全文:下载
  • 作者:Lee, Seong Ho ; Kim, Hae Kyu ; Park, Sung Chun
  • 期刊名称:Korean Journal of Anesthesiology
  • 印刷版ISSN:2005-6419
  • 出版年度:2010
  • 卷号:58
  • 期号:1
  • 页码:31-37
  • DOI:10.4097/kjae.2010.58.1.31
  • 语种:English
  • 出版社:The Korean Society of Anesthesiologists,
  • 摘要:Background

    We used warming fluid for maintenance of body temperature in operating room or intensive care unit. This study was aimed to investigate the effect of infusion rate and catheter length on the temperature of warming fluid.

    Methods

    Normal saline was used for testing infusion and temperature of infusion was maintained by a warmer as 40℃. The temperatures of solution in infusion line were measured at 0, 25, 50, 75, and 100 cm apart from warmer at six different flow rates (100, 200, 300, 700, 1,400, and 2,100 ml/h). We also measured the temperature changes at room temperature (RT) and 5℃, 10℃, and 15℃ above RT.

    Results

    The time to maintain solution temperature as 40℃ was 165, 122, 37, 37, 21, and 19 s at flow rate 100, 200, 300, 700, 1,400, and 2,100 ml/h. The peak temperature was 43.58 ± 0.58, 44.43 ± 1.18, 44.37 ± 0.70, 43.79 ± 0.61, 42.82 ± 0.97, and 42.11 ± 0.92℃ according to increasing flow rate. The temperature at 100 cm apart from warmer was 23.96 ± 1.53, 25.46 ± 2.76, 29.32 ± 3.47, 31.40 ± 5.38, 31.39 ± 6.75, and 38.14 ± 0.96℃ according to increasing flow rate.

    Conclusions

    These results suggested that the decreasing rate of temperature was related inversely to the flow rate and directly to the catheter length. There may be needed a rapid infusion pump with adequate heating system at a high flow rate and to locate the warmer close to patient for reserving a heating effect.

  • 关键词:Fluid temperature; infusion rate; Warmer