首页    期刊浏览 2024年07月18日 星期四
登录注册

文章基本信息

  • 标题:Fracture of J-tipped guidewire during central venous catheterization and its successful removal under fluoroscopic guidance -A case report-
  • 本地全文:下载
  • 作者:Park, Sun Kyung ; Yi, In-Kyong ; Lee, Jae-Ho
  • 期刊名称:Korean Journal of Anesthesiology
  • 印刷版ISSN:2005-6419
  • 出版年度:2012
  • 卷号:63
  • 期号:5
  • 页码:457-460
  • DOI:10.4097/kjae.2012.63.5.457
  • 语种:English
  • 出版社:The Korean Society of Anesthesiologists,
  • 摘要:

    Central venous catheterization by Seldinger's technique uses a guidewire which may cause complications such as kinking, knotting and fracture. Fractured guidewire may lead to severe outcomes such as embolization, and removal of it may also cause problems such as vessel damage. We experienced a case of right internal jugular venous catheterization complicated by guidewire fracture entrapped in the central venous catheter, and its successful removal under fluoroscopic guidance using snare-loop. The patient recovered without any complications. When resistance is felt during insertion or withdrawal of the guidewire, force should not be applied to the guidewire and care should also be exercised when passing the tissue dilator over the guidewire. Clinicians should be aware of this rare complication and snare-loop technique could be considered as one of the methods for removal of the fractured guidewire.

  • 关键词:central venous catheterization; Fracture; Guide wire