首页    期刊浏览 2024年10月05日 星期六
登录注册

文章基本信息

  • 标题:Determine a Proper Axial Length by Ultrasonic Biometry in Opaque Hydrophilic Acrylic Intraocular Lens
  • 本地全文:下载
  • 作者:Kim, Sang Hyuk ; Choi, Si Hwan
  • 期刊名称:Journal of the Korean Ophthalmological Society
  • 印刷版ISSN:0378-6471
  • 出版年度:2008
  • 卷号:49
  • 期号:9
  • 页码:1453-1460
  • DOI:10.3341/jkos.2008.49.9.1453
  • 语种:Korean
  • 出版社:The Korean Ophthalmological Society
  • 摘要:Purpose

    To investigate the accuracy of measuring ultrasonic axial lengths using A-scan (Ocuscan®, Alcon, USA) on opaque intraocular lenses after hydrophilic lens (ACRL-160®, Ophthalmed, USA) implantation.

    Methods

    We measured axial length through ultrasonic biometry prior to intraocular lens exchange. Twelve eyes of 14 patients who had intraocular lens opacity following hydrophilic acryl lens implantation were examined in a clinical study. We compared pre-cataract operative axial lengths to pre-IOL exchange axial lengths. The pre-IOL exchange axial lengths were measured in aphakic and pseudophakic modes. In the pseudophakic mode, the ultrasound velocity through an IOL was set at a rate ranging between 1,500 m/sec to 2,200 m/sec.

    Results

    The pre-IOL exchange axial lengths in the pseudophakic mode at the rate of 1,600 m/sec in lens velocity were the closest to pre-cataract operative values ( p =0.88).

    Conclusions

    When pre-cataract operative axial length is known with a hydrophilic acrylic intraocular lens, previous values may be used for IOL exchange of an opacity patient. If not, however, the closest values to pre-cataract operative axial lengths may be obtained by setting the velocity in the pseudophakic mode to a lens velocity rate of 1,600 m/sec.

  • 关键词:Axial lengths; Hydrophilic acryl lens; IOL opacity; Ultrasonic Biometry
国家哲学社会科学文献中心版权所有