  • 标题:Comparison of Colvardpupillometer, ORBScan II and Sirius in Determining Pupil Size for Refractive Surgery
  • 本地全文:下载
  • 作者:Baek, Ji Sun ; Park, Ji Hyun ; Yoo, Eun Seok
  • 期刊名称:Journal of the Korean Ophthalmological Society
  • 印刷版ISSN:0378-6471
  • 出版年度:2013
  • 卷号:54
  • 期号:8
  • 页码:1175-1179
  • DOI:10.3341/jkos.2013.54.8.1175
  • 语种:Korean
  • 出版社:The Korean Ophthalmological Society
  • 摘要:Purpose

    To compare pupil size measured by Colvard®pupillometer, ORBScan II®, Sirius® and analyze correlation between pupil size and night vision disturbance after refractive surgery according to measuring methods.

    Methods

    Two hundred eyes of one hundred patients who underwent refractive surgery were retrospectively evaluated. Pupil size was measured with Colvard®pupillometer, Sirius® in the scotopic light condition and with ORBScan II®, Sirius® in the photopic light condition. Patients filled out questionnaire about night vision disturbance at postoperative three months. Correlation between pupil size according to measuring methods and night vision disturbance was evaluated.

    Results

    The mean age of patient was 26 years and number of male patients was 36. In scotopic light condition, pupil size measured with Colvard®pupillometer (6.76 mm) was significantly larger than that of Sirius® (6.53 mm) ( p < 0.001). In photopic light condition, pupil size measured with ORBScan II® (3.98 mm) was significantly smaller than that of Sirius® (4.41 mm) ( p < 0.001). Night vision disturbance were correlated with pupil size and Sirius® in the scotopic light condition had strongest correlation among three measuring method (r = 0.44).

    Conclusions

    Sirius® tends to measure pupil size smaller than Colvard®pupillometer in the scotopic light condition. The correlation between night vision disturbance and pupil size was highest in Sirius®.

  • 关键词:Colvard®pupillometer; Night vision disturbance; ORBScan II®; Pupil size; Sirius®