首页    期刊浏览 2024年11月28日 星期四
登录注册

文章基本信息

  • 标题:The Analysis of Vault Change after Posterior Chamber Phakic Intraocular Lens Size Exchange
  • 本地全文:下载
  • 作者:Kim, Wook Kyum ; Cho, Eun Young ; Kim, Hee Sun
  • 期刊名称:Journal of the Korean Ophthalmological Society
  • 印刷版ISSN:0378-6471
  • 出版年度:2013
  • 卷号:54
  • 期号:11
  • 页码:1669-1674
  • DOI:10.3341/jkos.2013.54.11.1669
  • 语种:Korean
  • 出版社:The Korean Ophthalmological Society
  • 摘要:Purpose

    To evaluate the vault change after implantable collamer lens (ICL) size exchange according to the preoperative vault.

    Methods

    In 14 eyes of 13 patients, the vault change after ICL exchange operation due to unideal vault was compared in 2 groups, the smaller ICL exchanged group and larger ICL exchanged group.

    Results

    In 6 out of 14 eyes, the ICL was exchanged to a 0.5 mm smaller size and the vault was changed from 1.38 mm (1.18-1.70) to 0.71 mm (0.51-0.92) ( p = 0.03). In 8 eyes, the ICL was exchanged to a 0.5 mm bigger size and the vault was changed from 0.07 mm (0.03-0.13) to 0.50 mm (0.12-1.01) ( p < 0.01). The exchange operation was performed at 3.5 postoperative days (1-6) if the ICL was exchanged to 1 step smaller size, but the exchange operation was performed at 135 postoperative days (90-660) if the ICL was exchanged to 1 step bigger size ( p < 0.01).

    Conclusions

    ICL exchange to 1 step smaller or bigger size is an effective method to correct unideal postoperative vault to a more ideal vault size. The exchange to 1 step smaller size ICL tended to be performed sooner.

  • 关键词:Exchange; Implantable collamer lens; Phakic intraocular lens; Vault