  • 标题:Comparison of Optical Quality Between Two Intraocular Lenses Using Double-Pass Based Optical Quality Analysis System
  • 本地全文:下载
  • 作者:Kwak, Ae Young ; Choi, Moonjung ; Seo, Kyoung Yul
  • 期刊名称:Journal of the Korean Ophthalmological Society
  • 印刷版ISSN:0378-6471
  • 出版年度:2013
  • 卷号:54
  • 期号:12
  • 页码:1818-1823
  • DOI:10.3341/jkos.2013.54.12.1818
  • 语种:Korean
  • 出版社:The Korean Ophthalmological Society
  • 摘要:Purpose

    To compare postoperative optical qualities between two types of 1-piece aspheric intraocular lenses using the double-pass technique.

    Methods

    Uncorrected visual acuity (UCVA), best corrected visual acuity (BCVA), and spherical equivalent were evaluated in the subject groups; the first which included 25 eyes implanted with Tecnis® ZCB00 and the second which included 16 eyes implanted with Acrysof® IQ SN60WF. In addition, modulation transfer function (MTF) cut-off, Strehl ratio, and objective scattering index (OSI) were measured 6 months after cataract surgery in the 2 subject groups using Optical Quality Analysis System (OQAS, Visiometrics S.L., Terrasa, Barcelona, Spain) which is based on the double-pass technique.

    Results

    There were no significant differences in the spherical equivalent, UCVA, BCVA, and OSI between the 2 groups. However, both the MTF cut-off, and Strehl ratio showed statistically significant differences. The MTF cut-off (28.0 ± 7.79 vs. 20.4 ± 9.51 c/deg, p = 0.025) and Strehl ratio (0.14 ± 0.04 vs. 0.12 ± 0.05, p = 0.042) were higher in the Tecnis® ZCB00-implanted group.

    Conclusions

    The difference in characteristics of intraocular lenses subtly affects the vision quality as measured by values such as MTF cut-off and Strehl ratio after cataract surgery. OQAS based on the double-pass technique is considered useful in more objective estimates of the real retinal image quality after cataract surgery which is difficult to explain simply by measuring visual acuity.

  • 关键词:Double-pass; Optical Quality Analysis System; Visual quality; 1-piece aspheric intraocular lenses