  • 标题:Comparison of Intraocular Pressure Measured by Non-Contact Tonometer, Rebound Tonometer, Tono-Pen, and Goldmann Applanation Tonometer
  • 本地全文:下载
  • 作者:Lee, Kyung Sik ; Kim, Se Kyung ; Kim, Eung Kweon
  • 期刊名称:Journal of the Korean Ophthalmological Society
  • 印刷版ISSN:0378-6471
  • 出版年度:2014
  • 卷号:55
  • 期号:1
  • 页码:47-53
  • DOI:10.3341/jkos.2014.55.1.47
  • 语种:Korean
  • 出版社:The Korean Ophthalmological Society
  • 摘要:Purpose

    To compare the level of accuracy of intraocular pressure (IOP) measurements attained by non-contact tonometer (NCT), rebound tonometer (RT) Icare®, and Tono-Pen (TONO-PEN AVIA®), using Goldmann Applanation tonometer (GAT) as a reference value and to explore their clinical usefulness.

    Methods

    In a prospective study of 71 normal eyes, IOP was measured with NCT, RT, Tono-Pen and GAT. The IOP values of were then compared between the eyes.

    Results

    RT showed statistically most significant agreement with the GAT [ICC 0.811, 95%CI 0.712-0.878]. In analysis of Bland-Altman plots, NCT showed the smallest mean bias (+0.2 mm Hg) and widest CI (95%CI; ±5.05 mm Hg), RT showed relatively small mean bias (-0.7 mm Hg) and narrowest CI (95%CI; ±3.75 mm Hg).

    Conclusions

    There was a significant agreement between the RT and the GAT measurements. We expect RT to be considered as a reliable alternative when IOP measurement with GAT is not feasible.

  • 关键词:Goldmann applanation tonometer; Rebound tonometer