  • 标题:Utility of the Noncontact Specular Microscopy for Measurements of Central Corneal Thickness
  • 本地全文:下载
  • 作者:Yang, Young Seong ; Koh, Jae Woong
  • 期刊名称:Journal of the Korean Ophthalmological Society
  • 印刷版ISSN:0378-6471
  • 出版年度:2014
  • 卷号:55
  • 期号:1
  • 页码:59-65
  • DOI:10.3341/jkos.2014.55.1.59
  • 语种:Korean
  • 出版社:The Korean Ophthalmological Society
  • 摘要:Purpose

    To evaluate the efficacy of noncontact specular microscopy (NCSM) by comparing the measurement of central corneal thickness (CCT) to the measurement with anterior segment optical coherence tomography (AS-OCT) and ultrasound pachymetry (USP).

    Methods

    One examiner measured the CCT of 50 eyes of 50 healthy young subjects using NCSM, AS-OCT, and USP. The mean values and correlations were analyzed.

    Results

    The mean CCT value was 546.92 ± 32.06 µm with NCSM, 535.24 ± 30.54 µm with AS-OCT, and 546.38 ± 30.70 µm with USP. The CCT measurements with NCSM and USP were significantly thicker than with AS-OCT ( p < 0.001, p < 0.001, respectively). There were no significant differences between the measurements obtained with NCSM and USP ( p = 0.505). The 3 instruments were significantly correlated (r > 0.900 in all groups, p < 0.001 in all groups).

    Conclusions

    CCT measurements of healthy eyes using NCSM are more correlated with USP than AS-OCT. The CCT measurement using NCSM is a better alternative for USP than AS-OCT.

  • 关键词:Anterior segment optical coherence tomography; Central corneal thickness; Noncontact specular microscopy; Ultrasound pachymetry