To evaluation the accuracy of the IOL power calculation formulae measured by IOL Master® and applanation ultrasonography for the Tecnis ZCB00 IOL
MethodsWe performed a retrospective study of 170 eyes in 121 patients who underwent cataract surgery in our hospital with AMO Tecnis ZCB00 IOL.s. The SRK/T formula was used to predict the patient's implanted IOL power. Differences in the predicted refractive errors between IOL Master® and ultrasonography were analyzed and factors attributed to the differences were also analyzed. Three months after cataract surgery, mean numeric error and mean absolute error were analyzed.
ResultsSRK/II and SRK/T formulas calculated using ultrasonography showed differences compared to the same formulas calculated using IOL Master®, in which hyperopic shift was also demonstrated. No definite factor was attributed to the differences between the 2 methods. Although the 3 formulas of IOL Master® showed no significant difference in refractive errors, the SRK/T formula calculated using IOL Master® showed the least mean absolute and numeric errors.
ConclusionsIOL Master® is considered more suitable when determining proper AMO Tecnis ZCB00 IOL power in cataract surgery. The hyperopic shift should be considered when calculating the IOL power using only ultrasonography.