  • 标题:Comparison of Corneal Thickness and Anterior Chamber Depth Measured With Orbscan, Pentacam, and Ultrasound Pachymetry
  • 本地全文:下载
  • 作者:Park, Joo Youn ; Kim, So Young ; Jung, Moon-Sun
  • 期刊名称:Journal of the Korean Ophthalmological Society
  • 印刷版ISSN:0378-6471
  • 出版年度:2009
  • 卷号:50
  • 期号:5
  • 页码:664-669
  • DOI:10.3341/jkos.2009.50.5.664
  • 语种:Korean
  • 出版社:The Korean Ophthalmological Society
  • 摘要:Purpose

    We compared the measurements of corneal thickness and anterior chamber depth (ACD) using three different methods Orbscan, Pentacam and ultrasound pachymetry.

    Methods

    In healthy volunteers, central corneal thickness was measured with Orbscan, Pentacam and ultrasound pachymetry. Estimation of peripheral corneal thickness and ACD were done by Orbscan and Pentacam. All results were compared statistically.

    Results

    The mean central corneal thickness of 64 eyes measured by Orbscan, Pentacam and ultrasound pachymetry were 509.1±61.0 µm, 539.2±51.7 µm and 527.4±48.3 µm, respectively. There were statistically significant differences in the results among the three methods (repeated-measures analysis of variance (ANOVA), p <0.05). There were significant correlations between the ultrasonic pachymetry, the Orbscan and the Pentacam (Pearson correlation, r >0.90, p <0.05). The temporal and nasal peripheral corneal thicknesses were thinner in the Pentacam than in the Orbscan (paired t-test, p <0.05). The superior and inferior corneal thickness and ACD were not significantly different.

    Conclusions

    The results of the three instruments have significant correlations with each other, so all methods are clinically useful.

  • 关键词:Anterior chamber depth; Corneal thickness; Orbscan; Pentacam; Ultrasound pachymetry