首页    期刊浏览 2024年07月08日 星期一
登录注册

文章基本信息

  • 标题:Transcatheter closure of small ductus arteriosus with amplatzer vascular plug
  • 本地全文:下载
  • 作者:Cho, Eun Hyun ; Song, Jinyoung ; Kang, I-Seok
  • 期刊名称:Korean Journal of Pediatrics
  • 印刷版ISSN:1738-1061
  • 出版年度:2013
  • 卷号:56
  • 期号:9
  • 页码:396-400
  • DOI:10.3345/kjp.2013.56.9.396
  • 语种:English
  • 出版社:The Korean Pediatric Society
  • 摘要:Purpose

    The purpose of this study was to share our experience of transcatheter closure of small patent ductus arteriosus (PDA) by using an Amplatzer vascular plug (AVP).

    Methods

    We reviewed the medical records of 20 patients who underwent transcatheter closure at Samsung Medical Center and Sejong General Hospital from January 2008 to August 2012. The size and shape of the PDAs were evaluated by performing angiograms, and the PDA size and the AVP devices size were compared.

    Results

    The mean age of the patients was 54.9±45.7 months old. The PDAs were of type C (n=5), type D (n=12), and type E (n=3). The mean pulmonary end diameter of the PDA was 1.7±0.6 mm, and the aortic end diameter was 3.6±1.4 mm. The mean length was 7.3±1.8 mm. We used 3 types of AVP devices: AVP I (n=5), AVP II (n=7), and AVP IV (n=8). The ratio of AVP size to the pulmonary end diameter was 3.37±1.64, and AVP size/aortic end ratio was 1.72±0.97. The aortic end diameter was significantly larger in those cases repaired with AVP II than in the others ( P =0.002). The AVP size did not significantly correlate with the PDA size, but did correlate with smaller ratio of AVP size to aortic end diameter (1.10±0.31, P =0.032).

    Conclusion

    Transcatheter closure of small PDA with AVP devices yielded satisfactory outcome. AVP II was equally effective with smaller size of device, compared to others.

  • 关键词:Patent ductus arteriosus; Cardiac Catheterization; Vascular access devices