  • 标题:Microbial Contaminations of Ultrasonic Scalers, High-Speed Handpieces and Water Syringe Lines
  • 本地全文:下载
  • 作者:Chang, Beom-Seok ; Lee, Jung-Yun ; Han, Soo-Boo
  • 期刊名称:The Journal of the Korean Academy of Periodontology
  • 印刷版ISSN:0250-3352
  • 出版年度:1997
  • 卷号:27
  • 期号:4
  • 页码:941-947
  • DOI:10.5051/jkape.1997.27.4.941
  • 语种:Korean
  • 出版社:Korean Academy of Periodontology
  • 摘要:

    The purpose of this study was to evaluate various methods of decontamination of ultrasonic scalers, high-speed handpieces and air-water syringes in dental equipments.

    Eimination of possible sources of microbial infection in dental operatories should be of primary importance.

    Microbial contamination levels of high speed handpieces, air-water syringes and ultrasonic scalers of 11 dental units in Seoul National University Hospital were evaluated after flushing the lines, alcohol sponge rubbing, or soaking in 0.1% chlorhexidine for 1,3 and 5 minutes.

    The result suggests that flushing the lines or soaking the tips in 0.1% chlorhexidine before use of the water systems may reduce the microbial levels. Soaking in 0.1% chlorhexidine for 5 minutes was most effective in reducing bacterial contamination.

  • 关键词:Dental equipment; Microbial Infection; high speed handpiece, air-water syringe, ultrasonic scaler; flushing, alcohol, chlorhexidine