  • 标题:Effective dose from direct and indirect digital panoramic units
  • 本地全文:下载
  • 作者:Lee, Gun-Sun ; Kim, Jin-Soo ; Seo, Yo-Seob
  • 期刊名称:Imaging Science in Dentistry
  • 印刷版ISSN:2233-7822
  • 出版年度:2013
  • 卷号:43
  • 期号:2
  • 页码:77-84
  • DOI:10.5624/isd.2013.43.2.77
  • 语种:English
  • 出版社:Korean Academy of Oral and Maxillofacial Radiology
  • 摘要:Purpose

    This study aimed to provide comparative measurements of the effective dose from direct and indirect digital panoramic units according to phantoms and exposure parameters.

    Materials and Methods

    Dose measurements were carried out using a head phantom representing an average man (175 cm tall, 73.5 kg male) and a limbless whole body phantom representing an average woman (155 cm tall, 50 kg female). Lithium fluoride thermoluminescent dosimeter (TLD) chips were used for the dosimeter. Two direct and 2 indirect digital panoramic units were evaluated in this study. Effective doses were derived using 2007 International Commission on Radiological Protection (ICRP) recommendations.

    Results

    The effective doses of the 4 digital panoramic units ranged between 8.9 µSv and 37.8 µSv. By using the head phantom, the effective doses from the direct digital panoramic units (37.8 µSv, 27.6 µSv) were higher than those from the indirect units (8.9 µSv, 15.9 µSv). The same panoramic unit showed the difference in effective doses according to the gender of the phantom, numbers and locations of TLDs, and kVp.

    Conclusion

    To reasonably assess the radiation risk from various dental radiographic units, the effective doses should be obtained with the same numbers and locations of TLDs, and with standard hospital exposure. After that, it is necessary to survey the effective doses from various dental radiographic units according to the gender with the corresponding phantom.

  • 关键词:Radiography, Dental, Digital; Radiography, Panoramic; Radiation Dosage