出版社:Defence Scientific Information & Documentation Centre
摘要:RF receiver coils are very important parts of an NMR System. The design of these coils is very critical and has a dramatic effect on the SNR of the NMR signal and are generally developed in TRA/REC mode. This paper reports the developments of a 3.5 cm TRA/REC 26 MHz RF coil for P spectroscopy of small organs like thyroid. The coil is small in size, fits well in the neck for thyroid spectroscopy and is successfully working with the 1.5 tesla whole body Superconducting NMR System available at INMAS.
关键词:Thyroid spectroscopy;Superconductivity NMR System .